Quench and gas-mixing furnace
- Range
- vacuum to 1 bar; to 1,600 °C (p. 18)
- Precision
- about 1% in P; 1 to 15 °C (p. 18)
- Sample
- about 0.1 ml, hung on a Pt wire loop (pp. 19-20)
- Built from
- ceramic muffle tube, resistance windings, insulation, two gas flowmeters, a mixing chamber, Pt wire
- Original design
- Darken & Gurrey (1945); applied to magmas by Osborn (1959) (p. 19)
- Suited to
- liquidus relations in iron-bearing systems, mineral and glass synthesis, element partitioning (p. 20)
- In the book
- entry pp. 18-22; schematic Fig. 3.3, p. 21
What it does. Two feed gases, CO-CO₂ or H₂-CO₂, are mixed at set flow ratios and flowed over the sample, fixing ƒO₂ through the gas equilibrium (pp. 21-22, 172). A current pulse melts the fine quench wire and the sample drops into a bath, quenching in seconds, with nearly zero iron loss to containers and several samples per run (pp. 19-21).